![R&S NRP P Series Скачать руководство пользователя страница 65](http://html1.mh-extra.com/html/rands/nrp-p-series/nrp-p-series_user-manual_3112517065.webp)
Remote control commands
R&S
®
NRPxxP
65
User Manual 1179.5760.02 ─ 02
Parameters:
<window>
RECTangular | HAMMing | HANNing | BLACkman
*RST:
RECTangular
SENSe<Sensor>:TRACe:MPWidth?
Queries the attainable time resolution. The result is the smallest possible distance
between two pixels, i.e. it is the smallest time interval that can be assigned to a pixel.
Usage:
Query only
SENSe<Sensor>:TRACe:OFFSet:TIME
<time>
Adds an offset to the beginning of the trace sequence. Thus, the trace in the result dis-
play is moved in positive or negative x-direction. If you measure with more than one
power sensor, you can use this offset to arrange the traces to each other. The start of
recording relative to the trigger event is set using
Parameters:
<time>
Range:
Depends on the trigger delay.
*RST:
0.0
Default unit: Seconds
Example:
TRAC:OFFS:TIME 1.0
SENSe<Sensor>:TRACe:POINts
<points>
Sets the number of required values per trace sequence.
Parameters:
<points>
Range:
1 to 8192
*RST:
200
SENSe<Sensor>:TRACe:POINts:FPGA?
Queries the number of points used internally by the signal processing FPGA.
Usage:
Query only
SENSe<Sensor>:TRACe:TIME
<time>
Sets the trace length, time to be covered by the trace sequence. This time period is
divided into several equal intervals, in which the average power is determined. The
number of intervals equals the number of trace points, which is set using
.
Parameters:
<time>
Range:
10.0e-6 to 3.0
*RST:
0.01
Default unit: Seconds
Configuring the measurement modes