Remote control commands
R&S
®
NRPxxP
53
User Manual 1179.5760.02 ─ 02
Usage:
Event
TRIGger:LEVel
<level>
Effective only if
INTernal
.
Sets the trigger threshold for internal triggering derived from the test signal.
If an S-parameter device and/or the offset correction are enabled, the trigger threshold
is referenced to the correction data.
If the S-parameter device and/or the offset correction are disabled, the trigger thresh-
old and its input limits are adjusted as necessary.
If you enter a value without unit, the unit is defined by
Parameters:
<level>
Range:
10.0e-7 to 100.0e-3
*RST:
1.0e-4
Default unit: Watts
TRIGger:LEVel:UNIT
<unit>
Sets the unit of the trigger level if this value is entered without a unit. See also
Parameters:
<unit>
DBM | W | DBUV
*RST:
W
TRIGger:SENDer:PORT
<sender_port>
Selects the port where the power sensor outputs its own trigger event in case it is trig-
ger sender. See
for more information.
If the power sensor is the trigger sender, it can output its trigger event either on the
EXTernal<1>
or
EXTernal2
.
If the power sensor triggers itself, the trigger source of the power sensor must be
assigned to the other external port, as shown in the examples.
Parameters:
<sender_port>
EXT1 | EXTernal1 | EXT2 | EXTernal2
*RST:
EXT1
Example:
TRIG:SEND:PORT EXT1
TRIG:SOUR EXT2
TRIG:SEND:STAT ON
Example:
TRIG:SEND:PORT EXT2
TRIG:SOUR EXT1
TRIG:SEND:STAT ON