Remote control basics
R&S
®
NRPxxP
130
User Manual 1179.5760.02 ─ 02
Bit no.
Short description
Bit is set if
4
Critical
Critical static error exists. Status LED of the power sensor is
fast flashing red.
7
Legacy locked state
The power sensor is locked in the NRP legacy mode. Via the
SCPI channels (USBTMC or TCP/IP), only query commands
can be sent, but no setting commands.
When the first setting command is sent, the NRP legacy inter-
face takes precedence over all other command channels. This
bit is set to
1
, and all other channels can only execute query
commands. If a setting command is sent via a different chan-
nel, the power sensor indicates an error:
-200,"Execution error; sensor in LEGACY mode"
To leave the NRP legacy mode, close the NRP legacy chan-
nel. Either close the application which opened the NRP legacy
channel or close at least the connection to the power sensor.
10.2.3
Questionable status register
Contains information on questionable sensor states that occur if the power sensor is
not operated in compliance with its specifications.
0
0
0
Power summary
0
0
0
0
Calibration summary
POST failure (self test)
0
0
0
0
0
0
0
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
+
Figure 10-3: Questionable status register
Querying the register:
●
STATus:QUEStionable:CONDition?
●
Status reporting system