Instrument Functions
R&S
®
NGL200
65
User Manual 1178.8736.02 ─ 02.01
Table 6-5: File manager action
Action
Description
Copy from internal memory to USB.
Copy from USB to internal memory.
Delete the selected file.
6.9
Store and Recall
Upon power-up, the instrument loads the last stored settings from internal memory
location. Auto saved parameters are also applied during startup.
The R&S
NGL200 output states of all channels (Ch1, Ch2) are disabled when the
recall function is activated.
Auto saved instrument settings
Auto saved of instrument settings are applied when any of the following parameters
are changed:
●
Chapter 6.11, "General Instrument Settings"
●
USB connection mode
●
Ethernet settings
Inaddition of the auto saved instrument settings, the following instrument settings are
stored or recalled in the internal memory:
●
●
Settings in the Protection Function
●
●
●
●
1. Press [Settings] key.
The R&S
NGL200 displays the main menu window.
2. Select the "Device" tab to configure file settings for store and recall function.
The R&S
NGL200 displays the device menu.
3. Select "Save/Recall Device Settings".
The R&S
NGL200 loads the "Save/Recall Device Settings" dialog.
Store and Recall