Remote Control Commands
R&S
®
NGL200
124
User Manual 1178.8736.02 ─ 02.01
LOG:CHANnel[:STATe]
<state>
LOG:CHANnel[:STATe]?
Sets/Queries the data logging function for the previous selected channel on or off.
Parameters:
<state>
OFF | ON | 0 | 1
OFF | 0
Data logging function for the previous selected channel will be
deactivated.
ON | 1
Data logging function for the previous selected channel will be
activated.
*RST:
OFF | 0
Example:
LOG:CHAN ON
LOG:CHAN? -> 1
Data logging function for the previous selected channel is activa-
ted.
LOG:COUNt
<count>
LOG:COUNt?
[<count>]
Sets/Queries the number of measurement values to be captured.
Setting parameters:
<count>
numeric_value | MIN | MAX
numeric_value
Number of measurement values to be captured is set in the
range of 1 to 10000000.
MIN
Minimum number of measurement values to be captured is set
at 1.
MAX
Maximum number of measurement values to be captured is set
at 10000000.
Query parameters:
<count>
MIN | MAX
Return the number of measurement values.
Example:
LOG:COUN MAX
LOG:COUN? MAX -> 10000000
Data and File Management Commands