Remote Control Commands
R&S
®
NGL200
109
User Manual 1178.8736.02 ─ 02.01
[SOURce:]VOLTage:PROTection:MODE
<mode >
[SOURce:]VOLTage:PROTection:MODE?
Sets/Queries the OVP mode for the previously selected channel.
Parameters for setting and query:
<mode>
MEASured | PROTected
MEASured
The OVP turns off if the measured value exceeds the threshold.
PROTected
If the adjusted threshold is exceeded, you cannot turn on the
output of the instrument. In addition, the measured value is
monitored.
Example:
See
Example "Configuring the overvoltage protection"
[SOURce:]VOLTage:PROTection:TRIPped?
<state>
Queries the OVP state of the selected channel.
Example:
See
Example "Configuring the overvoltage protection"
VOLT:PROT:TRIP?
Response 1, the OVP is tripped.
Response 0, the OVP is not tripped.
Usage:
Query only
7.5.9
OPP Setting
The POWer:PROTection subsystem contains the commands for setting the overpower
protection parameters for the output channels. The default unit is W.
Configuration Commands