Remote commands to perform measurements in MSRA mode
R&S
®
FSWP-B1
75
User Manual 1177.5862.02 ─ 09
Example:
INST:SEL 'Spectrum2'
Selects the channel for "Spectrum2".
SYST:PRES:CHAN:EXEC
Restores the factory default settings to the "Spectrum2"channel.
Usage:
Event
Manual operation:
See
9.4
Configuring MSRA measurements
Configuring input/output and frontend settings
....................................................... 75
....................................................................................94
Adjusting settings automatically
..............................................................................99
9.4.1
Configuring input/output and frontend settings
The following commands are required to define input, output and frontend settings. Any
settings related to data acquisition or data output are only available for the MSRA pri-
mary application.
Configuring the vertical axis (amplitude, scaling)
....................................................79
...........................................................................................87
9.4.1.1
RF input
INPut<ip>:ATTenuation:PROTection:RESet
.........................................................................75
INPut<ip>:FILTer:HPASs[:STATe]
....................................................................................... 77
............................................................................................77
.........................................................................................78
INPut<ip>:LOSCillator:SOURce:EXTernal:LEVel
................................................................. 79
INPut<ip>:ATTenuation:PROTection:RESet
This command resets the attenuator and reconnects the RF input with the input mixer
for the R&S
FSWP after an overload condition occurred and the protection mechanism
intervened. The error status bit (bit 3 in the
STAT:QUES:POW
status register) and the
INPUT OVLD
message in the status bar are cleared.
The command works only if the overload condition has been eliminated first.