Remote commands to perform measurements in MSRA mode
R&S
®
FSWP-B1
77
User Manual 1177.5862.02 ─ 09
Suffix:
<ip>
.
1 | 2
irrelevant
Parameters:
<DirectPath>
AUTO | OFF
AUTO | 1
(Default) the direct path is used automatically for frequencies
close to 0
Hz.
OFF | 0
The analog mixer path is always used.
Example:
INP:DPAT OFF
Manual operation:
See
INPut<ip>:FILTer:HPASs[:STATe]
<State>
Activates an additional internal high-pass filter for RF input signals from 1
GHz to
3
GHz. This filter is used to remove the harmonics of the R&S
FSWP to measure the
harmonics for a DUT, for example.
This function requires an additional high-pass filter hardware option.
(Note: for RF input signals outside the specified range, the high-pass filter has no
effect. For signals with a frequency of approximately 4
GHz upwards, the harmonics
are suppressed sufficiently by the YIG-preselector, if available.)
Suffix:
<ip>
.
1 | 2
irrelevant
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
*RST:
0
Example:
INP:FILT:HPAS ON
Turns on the filter.
Manual operation:
See
INPut<ip>:FILTer:YIG[:STATe]
<State>
Enables or disables the YIG filter.
Suffix:
<ip>
.
1 | 2
irrelevant
Configuring MSRA measurements