Remote commands to perform measurements in MSRA mode
R&S
®
FSWP-B1
78
User Manual 1177.5862.02 ─ 09
Parameters:
<State>
ON | OFF | 0 | 1
*RST:
0
Example:
INP:FILT:YIG OFF
Deactivates the YIG-preselector.
Manual operation:
See
INPut<ip>:IMPedance
<Impedance>
This command selects the nominal input impedance of the RF input. In some applica-
tions, only 50
Ω
are supported.
Suffix:
<ip>
.
1 | 2
irrelevant
Parameters:
<Impedance>
50 | 75
numeric value
User-defined impedance from 50 Ohm to 100000000 Ohm
(=100 MOhm)
User-defined values are only available for the Spectrum applica-
tion, the I/Q Analyzer, and some optional applications.
*RST:
50 Ω
Default unit: OHM
Example:
INP:IMP 75
Manual operation:
See
INPut<ip>:LOSCillator:SOURce
<Location>
This command selects the type of local oscillator in the test setup.
Suffix:
<ip>
.
1..n
irrelevant
Parameters:
<Location>
EXTernal
External local oscillator connected to the "LO AUX Input" of the
R&S
FSWP.
INTernal
Internal local oscillator of the R&S
FSWP.
*RST:
INTernal
Example:
//Select external oscillator
CONF:PNO:MEAS RES
INP:LOSC:SOUR EXT
Configuring MSRA measurements