Remote Commands to Perform Measurements with I/Q Data
R&S
®
FSV/A3000 I/Q
Analyzer
193
User Manual 1178.8536.02 ─ 01
8.4.5
Configuring Data Acquisition
The following commands are required to capture data in the I/Q Analyzer.
Useful commands for I/Q data acquisition described elsewhere
●
●
[SENSe:]SWEep[:WINDow<n>]:POINts
●
Remote commands exclusive to I/Q data acquisition
....................................................................................... 193
.............................................................................................193
[SENSe:]IQ:BANDwidth:RESolution
.................................................................................194
...................................................................................... 194
.............................................................................................194
.................................................................................................195
...................................................................................195
................................................................................. 195
...................................................................................... 195
.............................................................................................199
...........................................................................................199
[SENSe:]IQ:BANDwidth:MODE
<Mode>
[SENSe:]IQ:BWIDth:MODE
<Mode>
This command defines how the resolution bandwidth is determined.
Parameters:
<Mode>
AUTO | MANual | FFT
AUTO
(Default) The RBW is determined automatically depending on
the sample rate and record length.
MANual
The user-defined RBW is used and the (FFT) window length
(and possibly the sample rate) are adapted accordingly. The
RBW is defined using the
command.
FFT
The RBW is determined by the FFT parameters.
*RST:
AUTO
Configuring I/Q Analyzer Measurements