Remote Commands to Perform Measurements with I/Q Data
R&S
®
FSV/A3000 I/Q
Analyzer
215
User Manual 1178.8536.02 ─ 01
In continuous sweep mode, the application calculates the moving average over the
average count.
In single sweep mode, the application stops the measurement and calculates the aver-
age after the average count has been reached.
Example:
SWE:COUN 64
Sets the number of sweeps to 64.
INIT:CONT OFF
Switches to single sweep mode.
INIT;*WAI
Starts a sweep and waits for its end.
Manual operation:
See
[SENSe:]SWEep:COUNt:CURRent?
This query returns the current number of started sweeps or measurements. This com-
mand is only available if a sweep count value is defined and the instrument is in single
sweep mode.
Return values:
<CurrentCount>
Example:
SWE:COUNt 64
Sets sweep count to 64
INIT:CONT OFF
Switches to single sweep mode
INIT
Starts a sweep (without waiting for the sweep end!)
SWE:COUN:CURR?
Queries the number of started sweeps
Usage:
Query only
[SENSe:]SWEep[:WINDow<n>]:POINts
This command defines the number of sweep points to analyze after a sweep.
Suffix:
<n>
.
Example:
SWE:POIN 251
Manual operation:
See
[SENSe:]SWEep:TIME
<Time>
This command defines the sweep time. It automatically decouples the time from any
other settings.
Capturing Data and Performing Sweeps