Remote Commands to Perform Measurements with I/Q Data
R&S
®
FSV/A3000 I/Q
Analyzer
196
User Manual 1178.8536.02 ─ 01
FLATtop
Flattop
GAUSsian
Gauss
RECTangular
Rectangular
P5
5-Term
*RST:
FLAT
Example:
IQ:FFT:WIND:TYPE GAUS
Manual operation:
See
[SENSe:]SWAPiq
<State>
This command defines whether or not the recorded I/Q pairs should be swapped (I<-
>Q) before being processed. Swapping I and Q inverts the sideband.
This is useful if the DUT interchanged the I and Q parts of the signal; then the
R&S
FSV/A can do the same to compensate for it.
Parameters:
<State>
ON | 1
I and Q signals are interchanged
Inverted sideband, Q+j*I
OFF | 0
I and Q signals are not interchanged
Normal sideband, I+j*Q
*RST:
0
Manual operation:
See
TRACe:IQ:BWIDth
This command defines or queries the bandwidth of the resampling filter.
The bandwidth of the resampling filter depends on the sample rate.
Parameters:
<Bandwidth>
Default unit: HZ
Manual operation:
See
TRACe:IQ:RLENgth
<NoOfSamples>
This command sets the record length for the acquired I/Q data.
Increasing the record length also increases the measurement time.
Note
: Alternatively, you can define the measurement time using the
SENS:SWE:TIME
command.
Configuring I/Q Analyzer Measurements