Remote Commands to Perform Measurements with I/Q Data
R&S
®
FSV/A3000 I/Q
Analyzer
194
User Manual 1178.8536.02 ─ 01
Example:
IQ:BAND:MODE MAN
Switches to manual RBW mode.
IQ:BAND:RES 120000
Sets the RBW to 120 kHz.
Manual operation:
See
[SENSe:]IQ:BANDwidth:RESolution
<Bandwidth>
[SENSe:]IQ:BWIDth:RESolution
<Bandwidth>
This command defines the resolution bandwidth manually if
is set to
MAN
.
Defines the resolution bandwidth. The available RBW values depend on the sample
rate and record length.
Parameters:
<Bandwidth>
refer to data sheet
*RST:
RBW: AUTO mode is used
Default unit: HZ
Example:
IQ:BAND:MODE MAN
Switches to manual RBW mode.
IQ:BAND:RES 120000
Sets the RBW to 120 kHz.
Manual operation:
See
[SENSe:]IQ:FFT:ALGorithm
<Method>
Defines the FFT calculation method.
Parameters:
<Method>
SINGle
One FFT is calculated for the entire record length; if the FFT
length is larger than the record length (see
), zeros are appended to the
captured data.
AVERage
Several overlapping FFTs are calculated for each record; the
results are averaged to determine the final FFT result for the
record.
The user-defined window length and window overlap are used
(see
and
*RST:
AVER
Example:
IQ:FFT:ALG SING
Manual operation:
See
Configuring I/Q Analyzer Measurements