Getting Started
R&S
®
FSPN
78
User Manual 1179.4363.02 ─ 01
4. In case of a
manual configuration
, define the resolution bandwidth and number of
cross-correlations for each half decade directly in the half decade configuration
table as absolute values.
The relative "RBW" factor and "XCORR Factor" are ignored.
3.4.4.3
DUT Power Supply
How to supply DUTs with current or voltage
Some measurement setups or DUTs require you to supply them with either voltage or
current.
The R&S
FSPN can provide both, when you set it up that way.
1. Enter the "Overview" dialog box and, from there, enter the "Output" dialog box.
2. Select the "DC Config" tab.
There you have the following options:
● Turn each of the three DC outputs on or off.
● Define the supplied voltage (or current in case of the V
supply
output).
● Define limits for the outputs. When you define limits, the supplied voltage will
not go below or above these.
●
3. Turn on the output of voltage or current with the "DC Power" feature.
Tip:
You can turn the output on and off easily with the "DC On / Off" button in the
toolbar. The "DC Config" button provides access to a softkey menu that lets you
define the signal characteristics without entering a dialog box.
Trying Out the Instrument