Remote Control
R&S
®
FSPN
394
User Manual 1179.4363.02 ─ 01
11.6.9.5
Y-Axis Scale
Commands to scale the y-axis in the frequency result display described elsewhere.
●
Start frequency:
Stop frequency:
Span:
●
[SENSe:]FREQuency:CENTer:STEP:LINK
●
[SENSe:]FREQuency:CENTer:STEP:LINK:FACTor
Commands to scale the y-axis in the phase result display described elsewhere.
●
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:AUTO
●
Auto scale once:
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:AUTO
CONFigure:TRANsient[:WINDow<n>]:AF:COUPling
..........................................................394
CONFigure:TRANsient[:WINDow<n>]:AF:COUPling:RPOint:MODE
.................................... 394
CONFigure:TRANsient[:WINDow<n>]:AF:COUPling:RPOint[:X]
..........................................395
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:MODE
.........................................................395
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:PDIVision
.................................................... 396
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:RPOSition
................................................... 396
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:RVALue
.......................................................397
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:UNIT
...........................................................397
CONFigure:TRANsient[:WINDow<n>]:AF:COUPling
<Mode>
This command selects the AF coupling mode.
Suffix:
<n>
.
1..n
Parameters:
<Mode>
AC
AC coupling: Frequency offset and phase offset are automati-
cally corrected and DC signal contains a phase offset of
±π
.
DC
DC coupling: Phase runs according to the existing frequency off-
set.
*RST:
AC
Example:
//Select AF coupling mode
CONF:TRAN:AF:COUP AC
Manual operation:
See
"Selecting the AF coupling mode"
CONFigure:TRANsient[:WINDow<n>]:AF:COUPling:RPOint:MODE
<Mode>
This command selects the zero phase positioning mode.
Common Measurement Settings