MEASure Subsystem
EM510
5.39
4065.7763.32-01.00
.
:BANDwidth:BETA<numeric_value>
Configuration of bandwidth measurement method Beta %.
Parameter:
<numeric_value>
0.1 to 99.9 %
*RST state:
1.0 %
Example:
MEASure:BANDwidth:BETA 10
.
:BANDwidth:BETA?
Query about the configuration of bandwidth measurement method Beta %.
Result:
%
value of bandwidth measurement method Beta %.
Example:
MEASure:BANDwidth:BETA? -> 10.0
.
:MODE CONTinuous|PERiodic
Setting the continuous or periodic measuring mode.
In the
PERiodic
measurement mode all detectors are discharged after the measuring time has elapsed,
and the next measurement is started. Only the individual measured values per measuring period are
displayed.
If the path to the result buffer
MTRACE
is enabled by command
TRACE:FEED:CONTrol MTRACE,
ALWays
, a measured value is stored in
MTRACE
each time the measuring time has elapsed.
In the
CONTinuous
measuring mode the measuring detector is read out every 200 msec, irrespective
of the measuring time. These current measured values are displayed.
The measuring time has an effect on the level detectors. With AVG the measuring time determines the
averaging time. With PEAK the measuring time determines the fall time. With FAST the measuring time
does not have any impact since it is only the current value which is measured.
If the path to the result buffer
MTRACE
is enabled by command
TRACE:FEED:CONTrol MTRACE,
ALWays
, a current measured value can be stored in
MTRACE
.
Parameters:
CONTinuous
continuous measurement
PERiodic
periodic measurement
*RST state:
CONTinuous
Example:
MEASure:MODE PERiodic
Содержание EM510
Страница 4: ......
Страница 6: ......
Страница 12: ...EM510 Contents 4065 7763 32 01 00 8...
Страница 17: ...Characteristics EM510 2 5 4065 7763 32 01 00 Fig 2 1 Front View Fig 2 2 Rear View...
Страница 18: ...EM510 Characteristics 4065 7763 32 01 00 2 6 Fig 2 3 Modules...
Страница 22: ...EM510 Characteristics 4065 7763 32 01 00 2 10 Fig 2 4 EM510 Detailed Block Diagram...
Страница 24: ...EM510 Characteristics 4065 7763 32 01 00 2 12...
Страница 30: ...EM510 Installation and Cabling 4065 7763 32 01 00 3 6...
Страница 180: ...EM510 Remote Control via LAN 4065 7763 32 01 00 5 144...
Страница 183: ...Maintenance and Troubleshooting EM510 6 3 4065 7763 32 01 00...