The selftest can be run with two different test routines. The basic test runs continuously in the
background and tests the test points inside the module. Based on this test, a "short test" or a "long test"
can be triggered. In the short test, a line spectrum is fed in at the antenna input and the receiver is set
to the line frequency nearest to the receive frequency. The complete receive path from the antenna
input of the tuner to the level evaluation is then measured and evaluated. In the long test, each line
frequency of the test spectrum is set and measured.
TEST? SHORt|LONG, REPort|QUIet
Triggering the "short test" or "long test".
Parameters:
SHORt|LONG
carry out short test | long test
REPort|QUIet
error messages in plain text are generated | not generated
Note:
If the test was initiated by
REPort
, plain text error reports are stored in the error queue. These reports
can be queried by SYStem:ERRor?.
Result:
0
no error detected
0
error detected :
Example:
TEST? LONG, QUIET -> 1
Содержание EM510
Страница 4: ......
Страница 6: ......
Страница 12: ...EM510 Contents 4065 7763 32 01 00 8...
Страница 17: ...Characteristics EM510 2 5 4065 7763 32 01 00 Fig 2 1 Front View Fig 2 2 Rear View...
Страница 18: ...EM510 Characteristics 4065 7763 32 01 00 2 6 Fig 2 3 Modules...
Страница 22: ...EM510 Characteristics 4065 7763 32 01 00 2 10 Fig 2 4 EM510 Detailed Block Diagram...
Страница 24: ...EM510 Characteristics 4065 7763 32 01 00 2 12...
Страница 30: ...EM510 Installation and Cabling 4065 7763 32 01 00 3 6...
Страница 180: ...EM510 Remote Control via LAN 4065 7763 32 01 00 5 144...
Страница 183: ...Maintenance and Troubleshooting EM510 6 3 4065 7763 32 01 00...