EM510
SENSe Subsystem
:MSCan
4065.7763.32-01.00
5.88
.
.
:DIRection?
Query about the scan direction.
Parameters:
none
Result:
UP, DOWN
Example:
MSCan:DIRection?
DOWN
.
.
:DWELl <numeric_value>|MINimum|MAXimum|INFinity
Setting the dwell time with the hold criterion fulfilled (T-DWELL in MSCAN-CONFIG menu).
Note:
According to the SCPI standard, this command is used to set the dwell time per scan step, i.e. the time
required by a step. This definition is met in the EM510 if the squelch is switched off. The hold criterion is
then fulfilled for each step.
Parameters:
<numeric_value>
dwell time in seconds
MINimum|MAXimum
lowest|highest dwell time
INFinity
infinite dwell time
*RST state:
0.5 s
Example:
SWEep:DWELl 10 ms
.
.
:DWELl? [MINimum|MAXimum]
Query about the dwell time.
Parameters:
none
query about the current dwell time
MINimum|MAXimum
query about the lowest|highest dwell time
Result:
Dwell
time in seconds; 9.9E37 is output for an infinite number
Example:
MSCan:DWELl? 0.010
Содержание EM510
Страница 4: ......
Страница 6: ......
Страница 12: ...EM510 Contents 4065 7763 32 01 00 8...
Страница 17: ...Characteristics EM510 2 5 4065 7763 32 01 00 Fig 2 1 Front View Fig 2 2 Rear View...
Страница 18: ...EM510 Characteristics 4065 7763 32 01 00 2 6 Fig 2 3 Modules...
Страница 22: ...EM510 Characteristics 4065 7763 32 01 00 2 10 Fig 2 4 EM510 Detailed Block Diagram...
Страница 24: ...EM510 Characteristics 4065 7763 32 01 00 2 12...
Страница 30: ...EM510 Installation and Cabling 4065 7763 32 01 00 3 6...
Страница 180: ...EM510 Remote Control via LAN 4065 7763 32 01 00 5 144...
Страница 183: ...Maintenance and Troubleshooting EM510 6 3 4065 7763 32 01 00...