CALibration Subsystem
4065.7763.32-01.00
5.22
.
: ROSCillator:DATE <year>,<month>,<day>
Setting the calibration date.
Result:
<year>,<month>,<day>
<year> =
1900..
<month> =
1..12
<day> =
1..31
Example:
CALibration:ROSCillator:DATE 2003,07,24
.
: ROSCillator:DATE?
Query about the calibration date.
Result:
<year>,<month>,<day>
<year> =
1900..
<month> =
1..12
<day> =
1..31
Example:
CALibration:ROSCillator:DATE? -> 2003,07,24
.
: ROSCillator:STORe
Storing the calibration value for setting the exact OCXO reference frequency and the calibration date.
Note:
When storing the calibration value and the calibration date the store command must be sent in one
single string. If the store command is sent without date an error message is generated:
1,"Device
dependent error;Calibration store only with date"
Example:
CALibration:ROSCillator:DATE 2003,07,24;STORe