4065.7763.32-01.00
5.100
Exemplifying all STATus register commands, for example the STATus:OPERation:SWEeping register,
the commands of the STATus:OPERation register are explained as follows.
STATus
.
OPERation
.
.
:CONDition?
Query about the condition section of the OPERation status register.
Parameters:
none
Result:
Depending on the setting by the
FORMat:SREGister
command, the contents of the register are
transferred as a decimal, binary or hexadecimal value in the ASCII code.
Example:
STATus:OPERation:CONDition? -> #H0008
.
.
:ENABle <numeric_value>
Setting the enable section of the OPERation status register.
Parameters:
<numeric_value>
value of the enable section (
0..65535 or #H0000..#HFFFF
or #B0..#B1111111111111111
)
*RST state:
will not be changed by *RST
Example:
STATus:OPERation:ENABle #H0008
.
.
:ENABle?
Query about the enable section of the OPERation status register.
Parameters:
none
Result:
Depending on the setting by the
FORMat:SREGister
command, the contents of the register are
transferred decimally, binary or hexadecimally in the ASCII code.
Example:
STATus:OPERation:ENABle? -> #H0008
Содержание EM510
Страница 4: ......
Страница 6: ......
Страница 12: ...EM510 Contents 4065 7763 32 01 00 8...
Страница 17: ...Characteristics EM510 2 5 4065 7763 32 01 00 Fig 2 1 Front View Fig 2 2 Rear View...
Страница 18: ...EM510 Characteristics 4065 7763 32 01 00 2 6 Fig 2 3 Modules...
Страница 22: ...EM510 Characteristics 4065 7763 32 01 00 2 10 Fig 2 4 EM510 Detailed Block Diagram...
Страница 24: ...EM510 Characteristics 4065 7763 32 01 00 2 12...
Страница 30: ...EM510 Installation and Cabling 4065 7763 32 01 00 3 6...
Страница 180: ...EM510 Remote Control via LAN 4065 7763 32 01 00 5 144...
Страница 183: ...Maintenance and Troubleshooting EM510 6 3 4065 7763 32 01 00...