Remote control commands
R&S
®
NRPxxP
83
User Manual 1179.5760.02 ─ 02
Parameters:
<magnitude>
0.0
≙
ideal matched source
1.0
≙
total reflection
Range:
0.0 to 1.0
*RST:
0.0
SENSe<Sensor>:SGAMma:PHASe
<phase>
Sets the phase angle of the complex reflection coefficient of the source,
Γ
source
.
Parameters:
<phase>
Range:
-360.0 to 360.0
*RST:
0.0
8.7.4.5
Using the S-Parameters program
The S-Parameters program helps loading an S-parameter table into the power sensor.
The S-Parameters program is part of the R&S
NRP Toolkit, see
To start the S-Parameters program
► In the Windows start menu, select "NRP Toolkit" > "S-Parameter Update Multi".
Configuring basic measurement parameters