![R&S 1444.1190K02 Скачать руководство пользователя страница 41](http://html1.mh-extra.com/html/rands/1444-1190k02/1444-1190k02_user-manual_774881041.webp)
Remote control commands
R&S
®
NRPxxP
41
User Manual 1179.5760.02 ─ 02
SENSe<Sensor>:AUXiliary
<mode>
Enables the measurement of additional-measured values that are determined together
with the main-measured value.
Parameters:
<mode>
NONE | MINMax | RNDMax
NONE
No additional values are measured.
MINMax
Minima and maxima of the trace are transmitted together with
the measured value.
Usually, extreme values are lost due to averaging the measured
values.
RNDMax
Not available in timeslot mode.
Randomly selected samples are transmitted. All evaluations use
these values instead of the average values.
*RST:
NONE
8.5
Controlling the measurement
The power sensor offers a bunch of possibilities to control the measurement:
●
Do you want to start the measurement immediately after the initiate command or
do you want to wait for a trigger event?
●
Do you want to start a single measurement cycle or a sequence of measurement
cycles?
●
Do you want to output each new average value as a measurement result or do you
want to bundle more measured values into one result?
8.5.1
Starting and ending a measurement
ABORt
Immediately interrupts the current measurement. If the measurement has been started
as a single measurement (
), the power sensor goes into the
idle state. However, if a continuous measurement is in progress (
ON
), the trigger system of the power sensor enters the waiting for trigger
state. When the trigger condition is met, a new measurement is immediately started.
See also
Chapter 8.5, "Controlling the measurement"
Controlling the measurement