![R&S 1444.1190K02 Скачать руководство пользователя страница 102](http://html1.mh-extra.com/html/rands/1444-1190k02/1444-1190k02_user-manual_774881102.webp)
Remote control commands
R&S
®
NRPxxP
102
User Manual 1179.5760.02 ─ 02
SENSe<Sensor>:TRACe:MEASurement:PULSe:SEParation?
Queries the time between two pulses. The time is measured between the first negative
edge and the subsequent positive edge of the pulse in seconds, where the edges
occur at crossings of the mid threshold. During this time, the pulse remains at the pulse
base level.
If the pulse period, and thus the duty cycle, cannot be determined, NaN for "not a num-
ber" is returned.
Usage:
Query only
SENSe<Sensor>:TRACe:MEASurement:TRANsition:NEGative:DURation?
Queries the time that the pulse requires to transition from the pulse top level to the
pulse base level. The value is read from the first falling edge in the analysis interval.
If the pulse period, and thus the duty cycle, cannot be determined, NaN for "not a num-
ber" is returned.
Usage:
Query only
SENSe<Sensor>:TRACe:MEASurement:TRANsition:NEGative:OCCurrence?
Queries the time when the signal passes through the medial reference power level with
falling edge, referenced to the delayed trigger event. Indicates the stop point of the first
power pulse within the analysis window.
Usage:
Query only
SENSe<Sensor>:TRACe:MEASurement:TRANsition:POSitive:DURation?
Queries the time that the pulse requires to transition from the pulse base level to the
pulse top level. The value is read from the first rising edge in the analysis interval.
Querying measurement results