![R&S 1444.1190K02 Скачать руководство пользователя страница 136](http://html1.mh-extra.com/html/rands/1444-1190k02/1444-1190k02_user-manual_774881136.webp)
Remote control basics
R&S
®
NRPxxP
136
User Manual 1179.5760.02 ─ 02
0
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
0
Sensor measuring
0
0
0
0
0
0
0
0
0
0
0
0
0
0
+
Figure 10-9: Operation measuring status register
Querying the register:
●
STATus:OPERation:MEASuring:CONDition?
●
STATus:OPERation:MEASuring[:SUMMary][:EVENt]?
Table 10-11: Used operation measuring status bits and their meaning
Bit no.
Short description
Bit is set if
1
Sensor measuring
Power sensor is measuring.
10.2.5.3
Operation trigger status register
The
CONDition
register contains information whether a power sensor is waiting for a
trigger event. The
EVENt
register contains information whether the power sensor has
been waiting for a trigger event since the last query.