Remote control commands
R&S
®
NRPxxP
82
User Manual 1179.5760.02 ─ 02
The complex reflection coefficient
Γ
sensor
of the power sensor, which is also required for
the correction, is prestored in the calibration data memory for many frequencies.
NRP
3-Path Diode Power Sensor
MHz to GHz, 100 pW to 200 mW (−70 dBm to +23 dBm)
SMART SENSOR TECHNOLOGY
Sensor
Source
Figure 8-8: Correction of interactions between the power sensor and the signal source
If the gamma correction is performed in combination with an S-parameter correction
(
SENSe<Sensor>:CORRection:SPDevice:STATe
ON
), the following is considered:
●
Interaction of the signal source with the S-parameter device
●
Input of the power sensor, depending on the transmission, expressed by the term
s
12
s
21
The interaction between the complex reflection coefficient
Γ
sensor
of the power sensor
and the reflection of port
2 is expressed by
s
22
rection is enabled, this interaction is always considered, regardless whether gamma
correction is performed or not.
Sensor
Source
S
21
S
12
S
22
S
11
NRP
3-Path Diode Power Sensor
MHz to GHz, 100 pW to 200 mW (−70 dBm to +23 dBm)
SMART SENSOR TECHNOLOGY
Figure 8-9: Correction of interactions between the power sensor, the signal source, and the S-param-
eter device
Remote commands:
SENSe<Sensor>:SGAMma:CORRection:STATe
................................................................. 82
SENSe<Sensor>:SGAMma:MAGNitude
............................................................................. 82
................................................................................... 83
SENSe<Sensor>:SGAMma:CORRection:STATe
<state>
Enables or disables the use of the complex reflection coefficient to correct the interac-
tions of power sensor and signal source.
Parameters:
<state>
*RST:
OFF
SENSe<Sensor>:SGAMma:MAGNitude
<magnitude>
Sets the magnitude of the complex reflection coefficient of the source,
Γ
source
.
Configuring basic measurement parameters