Page 3
SOF_7.5.1-07-04
Effective date: 28-Sep-07
CustomAir
®
by RAMVAC®
User Guide
INSTALLATION DATA
•
Indoor use only, in a dust free, climate controlled room.
•
DO NOT
install in an enclosed area where ambient temperature could exceed temperature specifications of
below 0º C / 35º F or above 40º C / 104º F while the compressor is running.
• Maintain minimum 6” clearance on sides, back, and top of all compressors. Maintain 12” clearance in front of compressor.
• On the 1029T compressor maintain a 12” clearance on left side for electrical service.
• Compressors are equipped with 4 adjustable rubber feet, to ensure unit stands firmly on the floor.
WARNING:
DO NOT
install on surfaces with more than a 5º incline.
WARNING:
To prevent tipping during transport securely fasten unit to shipping pallet.
CAUTION:
Danger of fire or explosion when using flamable substances.
CAUTION:
Never leave children unattended when compressor is in use.
•
Follow NEC, NFPA 99C, and all local codes.
•
Compressors are shipped with the appropriate electrical cord with plug.
•
Qualified personnel must install electrical outlets.
•
Electrical cicuit is a dedicated 20 amp circuit.
•
C1 Control and Motor Control Box for the 1029T compressor use a 20 amp time delay fuse.
•
See “Specifications” for more electrical information.
WARNING:
Electrical shock could occur as a result of improper grounding.
•
This product must be grounded according to NEC regulations and local codes.
WARNING:
Always
turn off compressor and remove power plug from outlet when servicing.
CAUTION:
Use of an extension cord is not advisable.
An undersized extension cord will cause
a drop in line voltage and loss of power. Overheating will result.
•
Attach drain tubing to fitting on bottom of coalescing filter and direct it into floor drain for disposal of water.
•
Connect compressor air outlet to facility plumbing with provided supply hose.
Note: Pressure regulator is customer supplied.
•
Connect Fresh Air Intake to external air source per NFPA 99C recommendations.
WARNING:
Property damage and / or personal injury may result if directions are not followed or
OEM parts are not used.
CAUTION:
Connect only equipment suitable for listed maximum pressure of the compressor.
If vital equipment is used, an alternative air source should be available in case of equipment failure.
Plumbing Connections
Electrical
Placement