Section II Installation
Electrical
•
Follow NEC, NFPA 99C, and all local codes.
•
Vacuum pump is shipped with the appropriate
6' electrical power conduit whip.
•
Qualified personnel must install a dedicated
electrical circuit.
•
See “Specifications” for more electrical
information.
Electrical shock could occur as a result of
improper grounding. This product must be
grounded according to NEC regulations and
local codes.
WARNING
Always turn off Vacuum pump and remove
power when servicing.
WARNING
Use of an extension cord is not advisable. An
undersized extension cord will cause a drop in line
voltage and loss of power. Overheating will result.
CAUTION
1.
2.
3.
F
G
H
E2 Control
Terminals
RAMVAC
Remote Control
Panel or
equivalent with
illuminated
24v Switch
F. G. H.
Switch
Light
Common
F. G
. H.
Electrician must provide a means to isolate the
circuit electrically from the supply mains on
all poles simultaneously. It shall be capable of
being locked in the OFF position.
WARNING
Vac Switch
Vac Light
Common
F - +24v DC
G - +24v DC
H - 24v DC Control
Illuminated Remote Switch
E2 Control
Electrols
Terminals
Toggle Switch
or other
non-illuminated
switch
(option)
F
H
E2 Control
Terminals
WARNING
This device should only be installed by licensed
electrical and plumbing contractors to all appli
-
cable federal, state or local building codes.
To avoid the risk of electrical shock, this
equipment must only be connected to a
supply mains with protective earth.
WARNING