MANUALE ISTRUZIONI / INSTRUCTIONS MANUAL
MTR605/025/DASM - R605/027/DASM - R605/170/DASM
Modello/Model R605DASM
H/ 1
e
lettr_
302_
ACSdhh
s.fm
H - E
LECTRICAL
POWER
CONNECTION
OF
THE
EXTERNAL
UNIT
TO
THE
GENER
-
AL
SYSTEM
:
WARNING
:
THE
SYSTEM
MUST
BE
SUPPLIED
BY
24VAC +/-
10%. A
NY
VOLTAGE
IN
EXCESS
WILL
DAMAGE
STEPPER
MOTOR
CONTROL
ELECTRONICS
WARNING
PRIOR
TO
CONNECTING
THE
COLLIMATOR
CHECK
THAT
SUPPLY
IS
PROTECTED
BY
AN
EXTERNAL
FUSE
6,3
A
(
SEE
THE
SPECIFICATIONS
).
CABLES
(24
VAC
)
USED
FOR
COLLIMATOR
CONNECTION
MUST
BE
SUITABLE
FOR
PURPOSE
AND
COLLIMATOR
CURRENT
ABSORPTION
.(
MIN
.
SECTION
1.5
MM
2)
THE
SYSTEM
REQUIRES
SCREENED
CABLES
WITH
THE
ADDITION
OF
FERRITE
CORES
TO
AVOID
INTERFERENCE
OF
ELECTROMAGNETIC
DISTURBANCES
WITH
THE
SERVER
ELECTRONICS
.
CABLES
FROM
THE
SYSTEM
TO
THE
EXTERNAL
BOARD
MUST
BE
SCREENED
.
THE
FERRITE
CORES
ARE
TO
BE
MOUNTED
ON
THE
CABLES
INSIDE
THE
BOX
AS
SHOWN
IN
THE
DRAWING
BELOW
.
FERRITE
CORES
ARE
USED
TO
ATTENUATE
INDUCTION
AND
IRRADIATION
OF
ELECTROMAGNETIC
INTERFERENCE
.
WE
SUGGEST
A
279
FERRITE
,
ATTENUATION
RANGE
100
MHZ
TO
200
MHZ
,
SUITED
TO
YOUR
CABLES
;
EG
:
MSFC
-10
FERRITE
BY
RICHCO
im
ages\collimat
o
ri\
a
cs\
ferrite.gif