MANUALE ISTRUZIONI / INSTRUCTIONS MANUAL
MTR605/025/DASM - R605/027/DASM - R605/170/DASM
Modello/Model R605DASM
C/ 4
ve
rifi
ca
_spec.fm
Validation of specification data:
• Validation of minimum filtration of the radiation unit (x-ray tube, collimator and pos-
sible filters) must be performed on a completely installed system by whoever is
responsible for the installation.
• Validation of light field luminosity and contrast is performed in-factory. See Chapter
COMPLIANCE
VERIFICATION
.
• Validation of system x-ray leakage is to be performed following the installation of
the system components under the responsibility of the person performing the instal-
lation.
• The correct installation of the system, electronic calibration and wiring layout of the
same are the responsibility of the engineer responsible for the installation of the
system.
• Protection of the cable is to be provided by the installation engineer.
Ralco is prepared to provide any information required regarding the validation methods
described above.
Classification
EN 60601-1 par. 5
• Protection against electric hazards: "Class I" equipment
• Protection against direct and indirect contacts: Type B equipment with applied
parts.
• Protection against water see page: "Common equipment"
• Safety of operation in the presence of inflammable anaesthetics with air or oxygen
or nitrous oxide: Equipment not suited to application in the presence of inflammable
anaesthetic mixtures containing air o oxygen of nitrous oxide.
• Operation conditions: Equipment for continuous operation at intermittent loads -
See Chapter
OPERATION INSTRUCTIONS
.
• Should label data on the collimator not correspond to the specifications herein,
inform Ralco of the non conformity.
• Verifications of the specifications are to be performed according to the indicated
equipment standards.
Operation environment:
• Ambient temperature = from 10°C to 40°C
• Relative Humidity = from 30% to 75%
• Atm. Press. = from 700 to 1060 hPa.
WARNING
GUIDES FOR ACCESSORIES PRIOR TO INSERTING ACCESORIES IN THE
GUIDES CHECK ON THE PERFECT FIT OF THE SUPPORT WITH THE MOUNTING
SLOT ON THE COLLIMATOR (TOLERANCE MAX. ± 0,5 MM). A FAULTY FIT COULD
BE DANGEROUS AND IT COULD CAUSE THE FALL OF ACCESSORIES