PROGRAMMING
4-24
4.7.3. Change LEL Gas Selection
This function allows selection of the measurement gas for the
LEL sensor from the on-board compound library. The
correction factor for this gas is automatically divided by the
correction factor for the span gas selected previously, to obtain
a new factor for the combination of gases. The new factor is
applied to the readings to display a true concentration. The user
can also create a custom correction factor for a non-standard
gas or mixture of gases.
1. If a combination LEL/VOL sensor is installed, pressing the
[Y/+]
key will cause the display to show:
LEL Gas =?
Methane
Otherwise, the message “No LEL installed” will be displayed
and you will return to the first sub-menu.
2. If the user does not want to change the LEL measurement
gas, press the
[Y/+]
key to accept the current gas and
skip
to Step 5.
3. To modify the LEL measurement gas, press the
[N/-]
key
first, then use the
[Y/+]
and
[N/-]
keys to scroll through a list
of gas names until the desired gas name appears in the
Display, then press the
[MODE]
key to select the new gas.
The display will read “Save new gas?” To confirm the new
gas, press the
[Y/+]
key. Press the
[N/-]
or
[MODE]
key to
discard the change and move to the next step.
Содержание QRAE PLUS PGM-2000
Страница 5: ...NOTE THE SYMBOL MEANS CAUTION RISK OF DANGER CONSULT OPERATION AND MAINTENANCE MANUAL ...
Страница 8: ......
Страница 12: ...GENERAL INFORMATION 1 4 non condensing ...
Страница 33: ...OPERATION 2 20 A new datalog event is created when the datalogging is resumed ...
Страница 34: ......
Страница 65: ......
Страница 67: ......
Страница 73: ......
Страница 81: ...TROUBLESHOOTING 8 8 ...
Страница 83: ...APPENDIX A A 2 ü Function allowed Password required ...
Страница 87: ...APPENDIX B B 3 Xylene p 2 8 1 1 Lower Explosion Limit of the gas in volume percent ...