PROGRAMMING
4-19
6)
Backlight Mode
– manual or automatic. If manual mode is
chosen, pressing the
[N/-]
key will turn on the LCD
backlight. Automatic mode allows the backlight to turn on
when the ambient light falls below a threshold level preset in
the monitor.
7)
Password
– a 4-digit numeric character which allows the
user to enter the programming mode when needed. The
manufacturer’s default password is “0000.”
8)
Pump Speed
– low or high. The factory default setting is
low, which reduces the flow rate by 30%, but increases the
battery run time by 10%, and increases the lifetime of the
LEL sensor. Setting the pump speed to high (see Section
4.7), offers quicker response time. A high pump speed is
required to measure reactive and slowly-responding gases
such as Cl
2
, PH
3
, NH
3
, and HCN.
9)
Averaging Method
– (TWA) Time Weighted Average
(default)
or (AVG) a running Average. These two averaging
methods are alternate ways to evaluate gas exposure over
time.
10)
Change Display Language
– allows the user to choose
either Spanish or English screen text.
11)
Temperature Units
– allows the user to display the
temperature in either degrees Centigrade or Fahrenheit.
12)
Enable/Disable Vibrator
– will activate or deactivate the
vibration alarm.
Содержание QRAE PLUS PGM-2000
Страница 5: ...NOTE THE SYMBOL MEANS CAUTION RISK OF DANGER CONSULT OPERATION AND MAINTENANCE MANUAL ...
Страница 8: ......
Страница 12: ...GENERAL INFORMATION 1 4 non condensing ...
Страница 33: ...OPERATION 2 20 A new datalog event is created when the datalogging is resumed ...
Страница 34: ......
Страница 65: ......
Страница 67: ......
Страница 73: ......
Страница 81: ...TROUBLESHOOTING 8 8 ...
Страница 83: ...APPENDIX A A 2 ü Function allowed Password required ...
Страница 87: ...APPENDIX B B 3 Xylene p 2 8 1 1 Lower Explosion Limit of the gas in volume percent ...