MultiRAE
User’s Guide
51
12.
In the menu that now appears on the left side, click “Policy Enforcement.” It is highlighted, and the
Policy Enforcement pane is shown:
For “Must Calibrate” and “Must Bump,” you have the options of no enforcement or enforcement
(including “Can’t Bypass,” and “Can Bypass”).
Must Calibrate.
The user is prompted to calibrate the instrument when calibration is due (as set by the
calibration interval). There are two programmable options:
Can’t Bypass.
Unless calibration is performed, the instrument cannot be used, and the only
option is to turn off the instrument.
Can Bypass.
If calibration is due but the user does not want to perform a calibration, the
instrument can still be used. In this case, the instrument records that the user has bypassed the
calibration requirement in a Policy Violation report.
Must Bump.
The user is prompted to bump test the instrument when a bump test is due (as set by the
bump test interval). There are two programmable options:
Can’t Bypass.
Unless a bump test is performed, the instrument cannot be used, and the only
option is to turn off the instrument.
Can Bypass.
If a bump test is due but the user does not want to perform one, the instrument can
still be used. In this case, the instrument records that the user has bypassed the bump testing
requirement in a Policy Violation report.
Содержание MultiRAE
Страница 1: ...MultiRAE User s Guide Rev D December 2013 P N M01 4003 000 ...
Страница 92: ......
Страница 93: ...Rev D December 2013 P N M01 4003 000 ...