- 48 -
F Shortcut Keys
You can declare which particular weighing function is to be launched using F1, F2, F3 or F4 keys.
For procedure informing you how to declare the particular function, read
F Shortcut Keys
section.
11.3.2. Setting the Reference Mass: Mass Determination from the Sample of Known Quantity
While determining the mass of a single piece,
ACAI
function (Automatic Accuracy Correction) is in
use.
Means of operation of ACAI function:
•
Number of pieces (on adding) on balance’s weighing pan has to be greater than before
•
Number of pieces (on adding) on balance’s weighing pan must be less than twice the amount
displayed before adding parts
•
Current quantity of parts must be within the
±
0,3 tolerance of the total value,
•
Measurement result has to be stabilised.
Procedure:
•
Place the container on the pan and tare its mass,
•
Press F button to which the function <
DETERMINE SAMPLE
> is assigned, wait to see the
editing window <
SAMPLE QUANTITY
>
•
Use arrow buttons
or
to select the correct sample quantity.
•
For optional quantity variable (displayed value <0000>) enter any number using arrow buttons
•
Confirm selected sample quantity, confirmation message - <
PLACExx PCS
> - shall be seen.
•
Place the declared number of pieces in the container and when the result is stable (the symbol
is displayed) confirm the mass by pressing
button,
•
The balance software automatically counts a single piece mass and enters the mode <
PARTS
COUNTING
> displaying the number of pieces which are on the pan (
pcs
). In the bottom line, a
single piece mass value is shown (if the option has been selected for <
INFORMATION
>
function).
CAUTION!
Remember that:
•
The total mass of all the pieces put on the weighing pan must not be greater than the balance maximum
weighing range.
•
The mass of a single piece cannot be lower than 0.1 reading unit of the balance. If this condition is not
met, the balance displays message: <Single part mass too low>.
•
While determining the number of pieces, wait for the stability pictogram
, next confirm the quantity.
Содержание PS 10100 R2