- 9 -
4. UNPACKING, ASSEMBLING AND STARTUP
4.1. Assembling and place of use
•
Balance should be stored and used in locations free of vibrations and shakes, free
of air movement and dust,
•
Ambient air temperature should not exceed the range of: +10
o
C
÷
+40
o
C
•
During balance operation, ambient temperature in the weighing room should not
change more than 3°C within one hour,
•
The balance should be located on a stable wall console desk or a stable working
table which is not affected by vibrations and distant from heat sources,
•
Take special safety measures when weighing magnetic objects, as part of the
balance is a strong magnet. Should such loads be weighed, use under hook
weighing option, which removes the weighed load from area influenced by the
balance’s magnet. For assembling the hook for under hook weighing option see the
bottom section of balance’s housing,
•
In order to avoid influence of static electricity on the measurement process, ground
the balance’s housing. The grounding bolt is located at the back of balance’s
housing.
Содержание PS 1000/X
Страница 2: ...2 APRILE 2013...