- 77 -
Fig. 65. Declaring printout content
−
On setting printout parameters go back to the weighing mode with procedure of
saving carried out changes.
Next, connect a label printer to balance’s communication interface using a dedicated
cable (see figure on cross-section of a connecting cable given in this user manual).
Check transmission parameters of a balance and printer – they should be equal.
Place a weighed load on balance’s weighing pan and on stabilization of measurement
result press PRINT key on balance’s overlay. The balance sends data to a connected
printer complying with saved template, and the printer prints 3 labels as designed in
the software.
Presentation of a label:
Fig. 66. A presentation of a printed label