- 82 -
Function
SEND MEASUREMENT RESULT IN CURRENT MEASURING UNIT
Command
SU CR LF sends result in current measuring unit on stabilization of
indication)
Function
IMMEDIATELY SEND MEASUREMENT RESULT IN CURRENT
MEASURING UNIT
Command
SUI CR LF
Function
ZERO BALANCE
Command
Z CR LF (zeroing of indication on stabilization)
Function
IMMEDIATELY ZERO BALANCE
Command
ZI CR LF
Function
TARE WHEN STABLE
Command
T CR LF
Function
IMMEDIATELY TARE BALANCE
Command
TI CR LF
Function
SWITCH OFF CONTINUOUS TRANSMISSION IN BASIC MEASURING
UNIT
Command
C0 CR LF
Function
SWITCH ON CONTINUOUS TRANSMISSION IN BASIC MEASURING
UNIT
Command
C1 CR LF
Function
SWITCH OFF CONTINUOUS TRANSMISSION IN CURRENT
MEASURING UNIT
Command
CU0 CR LF
Function
SWITCH ON CONTINUOUS TRANSMISSION IN CURRENT MEASURING
UNIT
Command
CU1 CR LF
Function
GIVE FACTORY NUMBER
Command
NB CR LF
Function
GIVE MEASURING RANGE
Command
FS CR LF
Function
GIVE SOFTWARE REVISION
Command
RV CR LF
Function
GIVE OR CHANGE DATE IN A BALANCE
Command
PD CR LF (causes sending data on set date or changes the date)
Function
GIVE OR CHANGE TIME IN A BALANCE
Command
PD CR LF (causes sending data on set time or changes the time)