- 65 -
Independently on the settings (YES / NO) on the printout of final result of statistical
calculations (enabled by pressing F key) comprises the following data.
N : 5
(number of measurements)
SUM : 161.121 g
(total mass of all statistical records in a measured series)
X : 32.224 g
(average mass of all statistical records in a measured series)
MIN : 20.486 g
(minimum mass)
MAX : 35.578 g
(maximum mass)
D : 15.092 g
(difference Max- Min)
SDV : 6.581 g
(standard deviation)
RDV : 20.4 %
(variance factor)
Fig. 55. Statistics – display content including a series of measurements
1. Mass currently placed on the weighing pan
2. Number of measurement in a measuring series
3. Sum of all completed measurements in a series
4. Average mass of all completed measurements in a series
5. Minimum saved mass in a carried out measuring series
6. Maximum saved mass in a carried out measuring series
7. The difference between the minimum and maximum saved mass in a measuring
series
8. Precisely calculated standard deviation value
9. Variance factor value
10. Measuring unit [g]
11. Working mode name
Statistics mode features a special option enabling quick calculating of statistics from a
measuring series with a possibility of generating reports from the completed calculations.
The user can declare the content of a header and footer in a printout of statistical report.
The header <PRINTOUT 1>, and the footer <PRINTOUT 2> are settable in non-standard
printouts.
In order to correctly use the option, the user has to create templates 1 and 2 using the non-
standard printouts (
see point
19.2
). As the templates are ready, the option can be enabled
in mode settings.