42
24.3. Chair Scales
Technical Specifications:
C315.K.250.C
Maximum capacity
250 kg
Readability [d=e]
100 g
Tare range
- 250 kg
Display
LCD (with backlight)
Communication interface *
RS 232, Wi-Fi
®
, USB
Power supply
100
240 VAC 50/60 Hz + battery
Operating temperature range
- 10
C ÷ + 40
C
Relative humidity
10
÷ 85 % RH, non-condensing conditions
*)
- conditioned by the scale type.
24.4. Bed Scales
Technical Specifications:
C315.4B.500.C
C315.8B.300.C
Maximum capacity
500 kg
300 kg
Readability [d=e]
200 g
100 g
Tare range
- 500 kg
- 300 kg
Display
LCD (with backlight)
Communication interface *
RS 232, Wi-Fi
®
, USB
Power supply
100
240 VAC 50/60 Hz + battery
Operating temperature range
- 10
C ÷ + 40
C
Relative humidity
10
÷ 85 % RH, non-condensing conditions
*)
- conditioned by the scale type.
25. TROUBLESHOOTING
Problem
Cause
Solution
Scale start-up fail.
Power supply disconnected.
Connect the power supply to the scale.
Battery discharged.
Connect the power supply to the mains,
charge the battery.
No battery (not installed or
installed incorrectly).
Check if the battery is installed correctly
(polarization).
The scale switches
off automatically.
<7.4.t1> parameter set to
value enforcing scale shut-
down after particular time
interval.
Go to <P7.Othr> menu, set <7.4.t1>
parameter to 'nonE' value.
During the start-up,
message
'LH'
is
displayed.
Weighing pan loaded during
the start-up.
Unload the weighing pan. Zero indication
is displayed.