23
Select a temporary unit and press
key, next go back to the home
screen.
12. SCALE PARAMETERS
Scale parameters are to be found in
<P2.rEAd>
submenu. The submenu
comprises functions allowing you to adjust your weighing device to ambient
conditions of a given workstation (filters) or individual needs (autozero on/off,
tare value memory).
12.1. Filter Level
Enter
<P2.rEAd / 2.1.FiL>
submenu.
Press
key, filter values are displayed successively one by one:
1
- Fast,
2
- Average,
3
– Slow.
Set a respective value and press
key to confirm, next go to the home
screen.
The higher the filter value, the longer the indication takes to
stabilise.
12.2. Value Release
Enter this parameter to adjust the rate of stabilisation of the measurement
result. Depending on the selected option, weighing time is either shorter
or longer.
Procedure:
Enter
<P2.rEAd / 2.2.APPr>
submenu.
Press
key, available values are displayed successively one by one:
F_P
- fast and reliable,
PrEc
– reliable,
FASt
– fast.
Press
key to confirm, next go to the home screen.