Example of user test voltages Us1 and Us2 change is given in Figure 2.2.
Figure 2.2 –
Menu. Modes
Voltages Us1 and Us2 can be also added to the list or removed from it at selection of
“Add” or “Delete” options respectively.
Option “TEST SETTINGS” (available only in the mode of insulation resistance
measurement, see item 2.3.5) the following is allowed:
- setting of insulation test duration from 1 to 10 minutes;
- selection of calculation formula for Dielectric Absorption Ratio: DAR
=R
60
/ R
15
or DAR
=R
60
/ R
30
;
- switching on/off of Polarization Index calculation.
2.3.2 Memory
In megohmmeter Е6-32, the memory is arranged as a set of 100 numbered objects,
each one consisting of 100 numbered cells. Additionally, the objects and the cells can be
assigned with their own names by means of RS-terminal program (see item 2.3.3).
After the measurement is finished, the display shows the last measurement result for 20
seconds. To save data, press button
, the result can be recorded to the selected cell of
selected object according to Figure 2.3.
11