3 MAINTENANCE AND TROUBLESHOOTING
3.1 General instructions
Maintenance includes meeting the rules of operation and storage.
Repair of megohmmeter is only allowed at the manufacturer
’
s site or in special repair
agencies.
3.2 Possible failures and troubleshooting procedures
Possible failures and troubleshooting procedure are provided in Table 3.
Table 3 – Possible failures and troubleshooting procedures
Failure symptoms
Possible cause
Troubleshooting procedure
Instrument fails to
switch on
Storage battery (battery) is
discharged
Check the storage battery
(battery) voltage, charge (see
item 2.2.2) or replace (see item
2.4), if necessary.
On Е6-32 there is no
image with
illumination switched
on
Failure of display contrast setting
for this temperature
Set the required display contrast
(see item 3.2.1).
Storage battery does
not charge
Failure of power unit or storage
battery
Check battery charger, if
necessary, replace storage
battery (see item 2.4)
The instrument does
not respond to
buttons pressing
Microprocessor failure
Switch off the instrument for 5
seconds and switch it on again.
Disconnect and then connect
the storage battery, if necessary
(see item 2.4).
3.2.1 Megohmmeter allows to change image contrast on the display without entering
the menu. For this press button
when the instrument is switched off, hold it and switch
the instrument on. Use buttons
and
to increase and decrease the contrast,
respectively. To finish the contrast setting press button
.
22