1.5 Design and Operation 

General view of  instrument is given in Figure 1.1.  

Figure 1.1  

1 - jack 

 for connection of the 

power unit (central pin 

– “

minus

); 

– 

protective plate (protective cover); 

– 

front panel; 

– 

handle (hook).  

6

Содержание E6-32

Страница 1: ... RESEARCH AND PRODUCTION COMPANY RADIO SERVICE JSC INSULATION RESISTANCE TESTER MEGOHMMETER Е6 32 OPERATION MANUAL ...

Страница 2: ... 84 to 106 kPa 630 to 795 mm Hg Protection grade of the casing IP54 in accordance with IEC 60529 Electrical safety as per IEC 61010 1 IEC 61010 2 30 2010 The instrument satisfy the EMC compatibility requirements complies with EMC directives complies with IEC 61326 1 2012 Due to permanent upgrade of instruments the items produced and the design described in this Operation Manual may differ to some ...

Страница 3: ...ulation to direct current Resistance measurement range Аccuracy 1 kΩ to 999 MΩ 3 3 digits 3 3 digits 1 00 to 9 99 GΩ 5 5 digits 250Vtestvoltages 10 0 to 99 9 GΩ from500Vtestvoltages 5 5 digits 100 to 300 GΩ from500Vtestvoltages 15 10 digits Resistance measurement range as per IEC 61557 2 from R Utest 1 mA to 10 GΩ at test voltage less than 500 V from R Utest 1 mA to 300 GΩ at test voltage more tha...

Страница 4: ...easurement units determining 1 3 3 Upon switching off the megohmmeters save the last measurement settings and restore them when switched on 1 3 4 Instrument stores up to 10000 measurement results with the possibility of data exchange with PC using specialized program RS terminal 1 3 5 Supply voltage range is from 7 5 to 5 2 V 7 Instrument provides self check of supply voltage At voltage decrease f...

Страница 5: ...livery set Description and designation Quantity 1 Insulation resistance tester megohmmeter Е6 32 1 2 Operation Manual 1 3 Power unit 1 4 Cable set including measuring cable RLPA 685551 002 red 1 5 m long connective cable RLPA 685641 002 black 1 5 m long connective cable RLPA 685641 002 01 blue 1 5 m long shielded measuring cable RAPM 685631 001 1 5 m long 1 1 1 1 5 Battery holder 5xAA RAPM 436244 ...

Страница 6: ...Design and Operation General view of instrument is given in Figure 1 1 Figure 1 1 1 jack for connection of the power unit central pin minus 2 protective plate protective cover 3 front panel 4 handle hook 6 ...

Страница 7: ...f the displayed information during insulation test see item 2 3 4 value of insulation resistance Dielectric Absorption Ratio Polarization Index After the active measurement is finished holding this button returns to the main menu In the menu mode the button allows scrolling down the menu 10 button is intended for modes switching selection of test voltage during insulation resistance measurement me...

Страница 8: ...t it shall be held under the operating temperature for two hours Check instrument for mechanical damages and dirt Check operability of protective covers and fasteners integrity of insulation and cleanness of cables Check the power unit for mechanical damages and dirt Check the date of last calibration of megohmmeter The calibration validity period shall not be expired Prior to instrument operation...

Страница 9: ...ularly for example by touch If the temperature grows fast stop charging 2 3 Operating the megohmmeter Е6 32 To switch the instrument on off press button After the instrument is switched on and self tested its software version is first shown on the display then the instrument switches over to mode of measurement that preceded switching off Supply voltage level is shown as miniature battery symbol i...

Страница 10: ...o the set of available test types fixed test voltages 50 100 250 500 1000 and 2500 V measurement of continuity Rcont measurement of varistor voltage Uvar measurement of spark gap breakdown voltage Udis as well as to change the values of user test voltages Us1 Us2 Active modes available at selection by button is marked by V sign see Figure 2 2 To add or remove the mode in the list of available ones...

Страница 11: ...on of calculation formula for Dielectric Absorption Ratio DAR R60 R15 or DAR R60 R30 switching on off of Polarization Index calculation 2 3 2 Memory In megohmmeter Е6 32 the memory is arranged as a set of 100 numbered objects each one consisting of 100 numbered cells Additionally the objects and the cells can be assigned with their own names by means of RS terminal program see item 2 3 3 After the...

Страница 12: ...LAST MEASURED RESULT option and save it in the selected cell of selected object according to Figure 2 3 To view the saved measurement results in the menu of instrument select MEMORY VIEW option the display shows the object selection window After selection the display shows information recorded to the current cell of the current object If there is no record in the selected cell then CELL EMPTY info...

Страница 13: ...description can be downloaded from the company s site www radio service ru For data exchange with a computer it is required turn on PC and run RS terminal program on the instrument which shall be switched on enter the mode Память Memory select option CONNECTION TO PC in the RS terminal window select all the necessary objects and cells A file with measurement results can be copied and edited by any...

Страница 14: ... Ratio DAR R60 R15 or DAR R60 R30 see item 2 3 6 switching on off of Polarization Index calculation see item 2 3 6 Example of parameters measurement is shown in Figure 2 4 Figure 2 4 Menu Measurement parameters Connection of megohmmeter Е6 32 for cable insulation resistance measurement is shown in Figures 2 5 and 2 6 а Measurement of resistances more than 10 GΩ with the rated accuracy shall be per...

Страница 15: ...gure 2 6 a In case of insulation test between transformer coils connect transformer case to the socket G see Figure 2 6 b Figure 2 6 Reduction of surface leakage currents influence Then use button to select the test voltage value By voltmeter readings check absence of residual or induced voltage at the object under test To start measurement press button twice Voltage generation is accompanied with...

Страница 16: ...ional information on Dielectric Absorption Ratio DAR and Polarization Index PI see item 2 3 6 press button To interrupt measurement before the set time press button Figure 2 7 Measurement of insulation resistance 16 ...

Страница 17: ...his case it is recommended to perform measurements twice with change of applied test voltage polarity This allows to determine the true value of insulation resistance as an average value of two measurements For short time measurements of the insulation resistance press and hold button the measurement is stopped after the button is released 2 3 6 Calculation of Dielectric Absorption Ratio and Polar...

Страница 18: ...60 R600 value of insulation resistance after 60 s and 600 s from the beginning of measurement respectively Polarization Index calculation is OFF on default it can be set to ON by the user during the measurement parameters setting see item 2 3 5 It is recommended to use the following factors KPOL to evaluate the insulation quality PI 1 service life of insulation is exhausted the process of insulati...

Страница 19: ...0 V Failure to comply with this rule can cause the instrument breakdown To start measurement press button The measurement results are shown on the display during 20 seconds and can be recorded to the instrument memory see item 2 3 2 Note that the true resistance of measurement object is lower than the instrument readings by the value of measuring cables resistance and transient resistances in poin...

Страница 20: ...on mode enter the correction menu select option Disable and press button Note If the resistance of the measurement object is much lower than that of the measuring cables results with a negative sign can be displayed because of measurement errors with enabled zero point correction 2 3 8 Measurement of varistor voltage discharge arresters breakdown voltage The modes are intended to check the complia...

Страница 21: ... to the instrument memory see item 2 3 2 2 4 Replacement of Storage Battery or Power Components Proceed as follows to replace the power components take out the screws that attach storage battery cover remove the cover and take the storage battery battery compartment out decouple the connector running to the storage battery battery compartment replace the storage battery or power components restore...

Страница 22: ...sary On Е6 32 there is no image with illumination switched on Failure of display contrast setting for this temperature Set the required display contrast see item 3 2 1 Storage battery does not charge Failure of power unit or storage battery Check battery charger if necessary replace storage battery see item 2 4 The instrument does not respond to buttons pressing Microprocessor failure Switch off t...

Страница 23: ...r humidity of 90 maximum at temperature of plus 30 ºС Exposure to atmospheric precipitations is not allowed 5 DISPOSAL Follow up local regulations regarding disposal of packages used batteries and accumulators Worn electrical and electronic equipment should be collected separately from the other waste it can not be placed together with the other waste Worn electronic equipment should be delivered ...

Страница 24: ...PRIMARY CALIBRATION Measurement device Insulation resistance tester megohmmeter Е6 32 Reg No ________________ on the basis of primary calibration results has been found compliant with the metrological requirements established in the type specification and fit for application in the field of measurement assurance state control Date of primary calibration ______________________ day month year Stamp ...

Страница 25: ...period from claim presentation till elimination of failures The guaranteed service life does not cover the storage battery To be filled in by the seller Date of sale ________________________________ Seller ________________________________________________________ Seller s address ________________________________________________ Seller s phone _________________________________________________ Stamp ...