12
RIDA qLine® Scan | User Manual
3.4 Environmental conditions
Place the instrument on a flat, horizontal surface with sufficient amount of space around
it. Avoid setting up at places where it can fall down.
Warning:
Please, pay attention that the pulled out drawer doesn’t reach out over the
working place. This might lead to injuries of the employees and can damage
the
instrument.
Avoid vibrations, strong electromagnetic waves (not computer), sources of heat and the
exposure to direct sunlight.
Keep the workplace surrounding the instrument free of any aggressive liquids.
3.5 Environmental parameters
Operation temperature: 15 °C to 35 °C
Working height: 0 to 2.000 m above sea level
Relative humidity: 20 % - 80 %, no condensing
Caution:
Before turning the instrument on, acclimatize the instrument at room temperature
to let all condensed water evaporate from the inside.
Содержание RIDA qLine Scan
Страница 1: ...User Manual RIDAqLine Scan https clinical r biopharm com...
Страница 2: ...2 RIDA qLine Scan User Manual...