1 Safety instructions

• 

The device is intended for use by professional users only.

• 

The guidelines for work in medical laboratories must be observed. 

• 

The operating instructions for the devices must be strictly observed.

• 

Please handle this equipment carefully at all times.

• 

Follow basic safety precautions to reduce the risk of injuries, fire and electrical shock.

• 

Read all information in this document attentively. Failures in understanding and 

  following the instructions may result in damage to the product, injury to operating    
  personnel and poor instrument performance.

• 

Electrostatic discharges can damage the device and affect the behavior and stability  

  of laboratory equipment. Therefore, avoid any electrostatic charge.

• 

Pay attention to all WARNING, CAUTION and STOP statements in this publication.

• 

Never open the electrical compartment while the equipment is plugged into  

  a power source.

• 

When using the RIDA qLine® Scan,  Good Laboratory Practice (GLP) must be observed. 

• 

Use a lab coat and pay attention to the safety standard procedures in you laboratory. 

• 

Please be careful while working with infectious material and pay attention to the safety  

  instructions thoroughly. Always wear gloves for your own safety. Proceed according to  
  the safety instructions of your laboratory when infectious substances get in contact  
  with your skin

  

Warning:

 

 

If the RIDA qLine® Scan or corresponding pieces are damaged, the system  

 

 

cannot work correctly and therefore the warranty will no longer be valid!

  Warning:
 

 

Always pay attention to the safety instructions while working with infectious  

  material

Содержание RIDA qLine Scan

Страница 1: ...User Manual RIDAqLine Scan https clinical r biopharm com...

Страница 2: ...2 RIDA qLine Scan User Manual...

Страница 3: ...es no liability for errors or omissions or for any damages resulting from the application or use of this information The content of this manual is the property of R Biopharm AG and is copyrighted Exce...

Страница 4: ...arameters 12 3 6 Transport and storage conditions 13 4 Installation and connection of RIDA qLine Scan 14 4 1 Transportation lock 14 4 2 Preparation 15 4 2 1 Connection 15 4 3 Installation 16 5 QC test...

Страница 5: ...instrument or training requirements if in any doubt before using the instrument Consult User Manual where symbol left is seen on the instrument Warnings Cautions and notes Special notes in these oper...

Страница 6: ...device Reference number Caution WEEE Waste Electrical and Electronic Equipment WEEE Waste Electrical and Electronic Equipment Refer to user manual Manufacturer Date of manufacturing RIDA qLine Scan s...

Страница 7: ...e avoid any electrostatic charge Pay attention to all WARNING CAUTION and STOP statements in this publication Never open the electrical compartment while the equipment is plugged into a power source W...

Страница 8: ...r professional use 2 2 Functions RIDA qLine Scan allows RIDA qLine Allergy tests to be scanned after manual processing The images are then quantitatively analyzed with RIDA qLine Soft Instrument featu...

Страница 9: ...ications General weight 2 9 kg Plastic contents 2 parts of body housing Drawer mechanical drawer Power Supply 1 x external power supply Mod 3A 183WP24 Input 100V 240V 50 60 Hz 0 6 A Output 24V 0 75 A...

Страница 10: ...tion This chapter describes how to install and start this device This includes unpacking accessories the software installation and PC connections Note Make sure you do not drop the scanner this could...

Страница 11: ...positioning of the device Place the instrument on a flat horizontal surface Remove all indicated safeguarding and adhesive tape Put no heavy objects onto the reader Warning Please pay attention to th...

Страница 12: ...es of the employees and can damage the instrument Avoid vibrations strong electromagnetic waves not computer sources of heat and the exposure to direct sunlight Keep the workplace surrounding the inst...

Страница 13: ...to 95 Storage Environmental temperature 20 C to 30 C Relative humidity 20 to 30 It is recommended to choose the transport and storage conditions in the medium range of the conditions specified above...

Страница 14: ...livered in a locked state The dark screw locks the scanning unit whereas the light colored screw seals a second opening In order to unlock the device please unscrew both screws completely remove the n...

Страница 15: ...ocket at the rear of the scanner Use only the external power adapter provided with the instrument Caution Only use the original power adapter of the RIDA qLine Scan The plugs cannot be mixed up due to...

Страница 16: ...The plugs fit into the dedicated sockets only and you cannot mix them up Please switch on the RIDA qLine Scan You will find the switch at the rear of the device The small blue lamp on the right front...

Страница 17: ...nnected PC To perform the QC test click the down arrow in Settings menu item and select Perform QC test Place 10 QC strips Art No ZG1108 in the strip holder included in the RIDA qLine Incubation Set I...

Страница 18: ...ow the results for each individual strip 6 Image creation with RIDA qLine Scan 6 1 Placing the test Pull out the drawer at the front side of the RIDA qLine Scan Open the drawer until you see the compl...

Страница 19: ...ning please check if the membranes were evaluated with good quality Repeat the scan if necessary 8 Turn off the reader At the rear of the device you can find a switch to turn the device on and off Whe...

Страница 20: ...de The cleaning of the interior is limited to the drawer area Warning Wear rubber gloves when cleaning the scanner Slight contaminations Pull out the drawer and then wipe the drawer with a damp cloth...

Страница 21: ...s procedure until the glass sheet is clean see picture Put the device back to its initial position carefully Please place the drawer back into its guideways and close it completely Finally reconnect t...

Страница 22: ...ers and end users in countries of the European Union EU must not dispose of electronic and electrical equipment as well as electronic and electrical accessories in household waste Within the EU please...

Страница 23: ...hat the RIDA qLine Scan is compliant with Regulation EU 2017 746 In Vitro Diagnostic Medical Devices In addition the RIDA qLine Scan fulfils the EMC protection requirements according to EN 61326 2 6 T...

Страница 24: ...24 RIDA qLine Scan User Manual R Biopharm An der neuen Bergstra e 17 64297 Darmstadt Germany E mail orders r biopharm de clinical r biopharm com Version 2022 04 0...