![R-Biopharm RIDA qLine Scan Скачать руководство пользователя страница 17](http://html1.mh-extra.com/html/r-biopharm/rida-qline-scan/rida-qline-scan_user-manual_3077678017.webp)
17
5 QC test using RIDA qLine® Soft
In order to check the perfect function of the RIDA qLine®
Soft it is recommended to use the RIDA qLine® QC kit
(ZG1108).
Start the RIDA qLine® Soft software on the connected PC.
To perform the QC test, click the down arrow in
Settings
menu item, and select
Perform QC test.
Place 10 QC strips (Art. No. ZG1108) in the strip holder,
included in the RIDA qLine® Incubation Set.
Insert it into the upper position (1) of the RIDA qLine® Scanner template. When using two
strip holders, you must simultaneously insert them into the (1) and (2) position in the
RIDA qLine® Scanner template.