MedTrack Rehabilitation Treadmills User Guide
Maintenance/Troubleshooting 3-7
Re-use
Before using the treadmill again after moving or storage,
check the power cord and all attachments to be sure they are
undamaged and securely connected, then test the system for
proper operation.
Troubleshooting
If the walking belt does not run, press the
Power
button twice.
This may eliminate the fault.
If problems persist, refer to the Troubleshooting Guide that
follows. Problems beyond the scope of this table may require
service assistance to isolate and correct. Contact Quinton
Technical Support for information: 800-426-0337.
Warning! Do not remove the treadmill hood: Dangerous
voltages are present. There are no operator-
serviceable components.
♦
Servicing should be done only by qualified service personnel
who should consult the service manual before attempting any in-
depth troubleshooting.
Error Codes
The controller displays an error message (the letter E
followed by three digits) if it detects a problem in the system.
If an error occurs during operation, the treadmill decelerates
and the walking belt stops. Only the E201 error code permits
the treadmill to keep running. The error code remains on the
display until the power is turned off. If an error code appears,
record the code number, then recycle the power. If the error
persists, call Quinton Technical Support.