microPATCalibration 

 

Your microPAT is calibrated for twelve months from the 
date of purchase. 
 
Have your microPAT re-calibrated with Quick Test any 
time during the twelfth month and the calibration will be 
valid for a full twelve months from the due date. 
 
 
Contact [email protected] for details. 

PAT Test Training 

 

The Quick Test training team have successfully trained 
thousands of people with all levels of experience. 
 
We provide training in a professional and friendly 
manner. 
 
Contact [email protected] for details 

Содержание microPAT

Страница 1: ...microPAT user manual...

Страница 2: ...welfth month and the calibration will be valid for a full twelve months from the due date Contact calibration quick test biz for details PAT Test Training The Quick Test training team have successfull...

Страница 3: ...lts Display 5 Switched On Indicator 5 Tests 6 Test Table 6 Identifying Class 1 and Class 2 7 Testing a Class 1 Appliance 7 Testing a Class 2 Appliance 8 Testing an Extension Lead 9 Testing an IEC Lead...

Страница 4: ...for surge protected extension leads Tests different lengths of extension leads Appliance switched on check User calibration check In service Inspection Testing Always carry out a full visual inspectio...

Страница 5: ...3 Overview Charger Connector Results Indicators Test Buttons Battery Charge Indicator IEC Connector Test Lead Connector Switched On Indicator Appliance Socket...

Страница 6: ...e plugged into the charger Battery Indicator The Battery Charging indicator will flash green for the first hour when connected to the charger This is a fast charge period After this the indicator will...

Страница 7: ...rresponding red FAIL indicator will light up Additionally the indicator next to the test button will be green or red A new test can be started as soon as the results are being displayed there is no ne...

Страница 8: ...al connections This is then used to measure the electrical resistance of the appliance to earth or any exposed metal part This resistance has to be very high to protect the user Insulation Resistance...

Страница 9: ...ible to switch the appliance on press the Class 1 button again to proceed with the test The Class 1 test button will flash amber while the test is underway When the test is finished either the green P...

Страница 10: ...e Switched on indicator will flash to prompt you to do so If the appliance is then switched on the tester will proceed with the test If it is not possible to switch the appliance on press the Class 2...

Страница 11: ...flash amber while the test is underway When the test is finished either the green PASS indicator will light up or the appropriate red FAIL indicators will light up The test button will also be green...

Страница 12: ...test instrument should be verified and recorded annually or in accordance with the manufacturer s instructions The microPAT should be formally calibrated annually Please contact Quick Test to arrange...

Страница 13: ...ipment 110V and 3 Phase equipment is tested in exactly the same way as any other equipment The only additional requirement is the use of appropriate adapters to connect to the microPAT Please contact...

Страница 14: ...1mA Tolerance 5 Limits Class 1 Cable 2M Class 2 4M Supply Voltage 12V 5 Current 400mA max Regulatory Information LVD EN 61010 EMC EN 55011 12 The microPAT is produced by a registered WEEE producer Reg...

Страница 15: ...tion Quick Test warrants your microPAT against defects in materials and workmanship for a period of one year from the date of purchase This warranty covers only those defects which arise as a result o...

Страница 16: ...www quick test biz...