Model QS-500 Tubestand
30
Chapter 3 Assembly & Installation
table first, then the vertical wall stand. Alignment instructions are provided in
the installation and operation manual provided with each component.
Refer to the collimator manual for additional alignments and performance verifi-
cations associated with beam limiting devices.
Collimator Light Field Versus Actual X-Ray Field
1. Using any size film with cassette, set the proper technique in the x-ray gen-
erator.
2. Make an exposure and develop the film to verify actual x-ray field versus
collimator light field alignment is within required specifications.
3. Use collimator horizontal cross light to align the table. Rotate the x-ray
tube to check the pitch adjustment as required.
4. Verify vertical movement and alignment with bucky stand or cassette holder
using collimator light.
Содержание QS-500 Series
Страница 4: ...Model QS 500 Tubestand THIS PAGE INTENTIONALLY LEFT BLANK ii Revision History...
Страница 6: ...Model QS 500 Tubestand Table of Contents THIS PAGE INTENTIONALLY LEFT BLANK iv...
Страница 12: ...Model QS 500 Tubestand THIS PAGE INTENTIONALLY LEFT BLANK Warranty Information x...
Страница 13: ...Chapter INTRODUCTION 1 1...
Страница 14: ...2...
Страница 17: ...5 Chapter SPECIFICATIONS 2...
Страница 18: ...6...
Страница 22: ...Model QS 500 Tubestand 10 Chapter 2 Specifications Figure 2 Model QS 500 Tubestand Dimensions...
Страница 23: ...11 Chapter ASSEMBLY INSTALLATION 3...
Страница 24: ...12...
Страница 43: ...Chapter OPERATION 4 31...
Страница 44: ...32...
Страница 48: ...Model QS 500 Tubestand Chapter 4 Operation 36 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 49: ...Chapter MAINTENANCE 5 37...
Страница 50: ...38...
Страница 56: ...Model QS 500 Tubestand Chapter 5 Maintenance 44 Figure 26 Model QS 500 Tubestand Parts Location Diagram 9...