Model QS-500 Tubestand
Vertical Column Angle Adjustment
1. Place a double-bubble level, two feet in length minimum, on the front of the
Vertical Column (see Figure 8).
2. If not level, adjustment can be made by moving the Floor Track in or out
from wall. Ensure the Vertical Column is level throughout the full length of
longitudinal travel.
3. Secure the Floor Track to floor using appropriate fastening devices
(depending on floor construction), each capable of supporting 200 lbs. pull-
load.
19
Chapter 3 Assembly & Installation
Figure 8. Checking the Tubestand Vertical Angle
VERTICAL
COLUMN
Содержание QS-500 Series
Страница 4: ...Model QS 500 Tubestand THIS PAGE INTENTIONALLY LEFT BLANK ii Revision History...
Страница 6: ...Model QS 500 Tubestand Table of Contents THIS PAGE INTENTIONALLY LEFT BLANK iv...
Страница 12: ...Model QS 500 Tubestand THIS PAGE INTENTIONALLY LEFT BLANK Warranty Information x...
Страница 13: ...Chapter INTRODUCTION 1 1...
Страница 14: ...2...
Страница 17: ...5 Chapter SPECIFICATIONS 2...
Страница 18: ...6...
Страница 22: ...Model QS 500 Tubestand 10 Chapter 2 Specifications Figure 2 Model QS 500 Tubestand Dimensions...
Страница 23: ...11 Chapter ASSEMBLY INSTALLATION 3...
Страница 24: ...12...
Страница 43: ...Chapter OPERATION 4 31...
Страница 44: ...32...
Страница 48: ...Model QS 500 Tubestand Chapter 4 Operation 36 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 49: ...Chapter MAINTENANCE 5 37...
Страница 50: ...38...
Страница 56: ...Model QS 500 Tubestand Chapter 5 Maintenance 44 Figure 26 Model QS 500 Tubestand Parts Location Diagram 9...