7.  Operating Instructions 

CAUTION

 

  Do not operate the power supply unless it is connected to the converter. 

  Never allow liquid to spill into the converter.    

  Do not allow a microtip to vibrate in air for more than 10 seconds.  

  Do not allow the vibrating microtip to contact anything but the sample. 

  Never place a washer between the probe and the converter. 

  Never apply grease to the mating surfaces or threads of the converter, probe or microtip. 

  Should it become necessary to remove a probe, use the wrenches supplied.

 

Never 

attempt to remove the probe by twisting the converter housing, as this may damage the 
electrical connections within the housing. 

 
 
 

Set-up: 

1.  Plug the electrical line cord into the electrical outlet. If the unit is already on, depress the 

OFF key. 

2.  If the optional foot switch is used, insert the plug into the jack located on the rear panel. 
3.  If the convertor and probe assembly is not already assembled; screw the probe onto the 

convertor and using the wrenches provided tighten it securely. (See instruction in 
Maintenance section for attaching and detaching microtips). 

4.  If using a laboratory stand, mount the convertor /probe assembly using a clamp.  Be sure to 

secure the clamp to the upper section of the convertor housing only.  Do not secure the 
clamp to any other portion of the convertor/probe assembly. 

5.  Connect the converter cable to the power supply. 

 

 

Operation: 

Press the ON key. The screen will display the power rating of the Ultrasonic Processor and the 
following control parameters. 

 

 

 

 
 
 
AMPLITUDE:

 Desired amplitude must be set in order for the Ultrasonic Processor to be 

operational. The other control parameters – Time and Pulse, do not have to be set for 
continuous operation.  AMPL displays the amplitude selected e.g. 40%. To set the amplitude at 
40% when the ultrasonics is off, press the AMPL key and the numeric keys for a 40% reading 
on the screen, and then press the ENTER/REVIEW key.   

Note: The minimum amplitude setting is 20%. 

 

The screen will display:

 

 

 

 

Содержание Sonicator

Страница 1: ...11 Sonicator ULTRASONIC PROCESSOR Part No Q125 OPERATION MANUAL 53 Church Hill Road Newtown CT 06470 U S A Phone 203 426 0101 1 877 338 9636 Fax 203 426 7026 Web www sonicator com E mail info sonicato...

Страница 2: ...Rev 3 11 2...

Страница 3: ...pecifications 6 7 4 Principles of Operation 8 9 5 Description of Components Functions of Controls 10 12 6 Preparation for Use 13 7 Operating Instructions 14 17 8 Maintenance 18 20 9 Troubleshooting 20...

Страница 4: ...rranty of either merchantability or fitness for a particular purpose No person or company is authorized to change modify or amend the terms of this warranty in any manner or fashion whatsoever Under n...

Страница 5: ...Generator within a Fume Hood Hearing protection is highly recommended It is recommended that a sound abating enclosure or ear protection be used when operating the Ultrasonic Processor NEVER immerse...

Страница 6: ...slo blo 1 6 Amps slo blo Weight 7 lbs 3 2 Kg Dimensions 8 W x13 75 L x 5 75 H 203mm x 349mm x 146mm Output Voltage 1000 V rms max Output Frequency 20 KHz Converter Weight 0 75 lbs 0 34Kg Dimensions 6...

Страница 7: ...e used If the 220V plug is not configured to match the wall receptacle a properly grounded universal AC socket adapter must be added Important Universal adapters do not convert voltage or frequency Fi...

Страница 8: ...ore information on this subject Please consult with a product specialist for assistance with selecting a probe for your application Relationship of Amplitude and Wattage Sonication power is measured i...

Страница 9: ...rvation the amount of power required cannot be predetermined A sensing network continuously monitors the output requirements and automatically adjusts the power to maintain the amplitude at the presel...

Страница 10: ...f Controls The Model Q125 includes a standard 1 8 diameter microtip 4422 Q125 Front Panel Q125 Rear Panel Generator Converter 1 8 Microtip Converter Cable Converter Cable Connector Power Cord Connecto...

Страница 11: ...main power on 0 key Switches the main power off AMPL Controls the amplitude of vibration at the probe tip key Used with the AMPL key when the unit is on stand by to set the amplitude of vibration at...

Страница 12: ...to pin 6 9 PIN D SUB CONNECTOR Pin No Description 1 Not connected 2 Not connected 3 Not connected 4 Enables connection to a frequency counter 5 Enables connection to an external power monitor 5 mv 1...

Страница 13: ...sary to convert the unit for different voltage operation proceed as follows 1 Ensure that the power cord is not connected to the electrical outlet 2 Open the fuse holder cover using a small screwdrive...

Страница 14: ...the convertor and probe assembly is not already assembled screw the probe onto the convertor and using the wrenches provided tighten it securely See instruction in Maintenance section for attaching an...

Страница 15: ...ative If the process is initiated by the footswitch the STOP key becomes inoperative 3 To increase or decrease the amplitude in small increments when the ultrasonics is on depress the AMPL to display...

Страница 16: ...et independently from 01 second to 59 seconds During the OFF portion of the cycle the red indicator on the PULSE key will illuminate If the OFF portion of the cycle exceeds three seconds a cautionary...

Страница 17: ...processing time and elapsed processing time e g Set 0 30 00 Time 0 22 10 c Selected pulsing cycle and actual pulsing cycle e g Pulse 01 01 01 00 d Amount of power in watts and accumulated amount of en...

Страница 18: ...nance Ultrasonic processors create high intensity vibration which puts stress on the converter and horn assembly The sides and end of the probe must never be allowed to come in contact with anything b...

Страница 19: ...cotton swab to remove any debris on the threading of the connecting stud Allow the alcohol to dry completely 3 Clean threading in converter Use alcohol and a cotton swab to remove any debris on the th...

Страница 20: ...otip being operated is damaged or worn past its useful life A fuse s has failed If a fuse s has failed proceed as follows 1 Turn the unit off by depressing the OFF key and disconnect the line cord fro...

Страница 21: ...ion Number prior to returning the instrument Care should be exercised to provide adequate packing to insure against possible damage in shipment The Ultrasonic Processor should be sent to the Service D...

Страница 22: ...h or exposed to radiological biological or chemical agents and has been decontaminated rendering it safer for handling use or disposal Authorization By accepting authorization to return the equipment...

Отзывы: